INSTALLATION
BIC-II
The BIC-II should be mounted in an area that will allow operation within the specified operating temperature
range and that is suitable for NEMA 4 enclosures and polycarbonate material. The front cover should be
accessible and have sufficient room to swing open.
Refer to Fig. 1 - Outline and Mounting Detail.
It is advisable to keep the BIC-II away from high voltage or current runs, SCR control drives and RF
interference.
Do not mount the BIC-II in direct sunlight without the use of a sun shield.
INTERCONNECTION
Communication cables may be run in a common grounded metal conduit separate from power wiring.
Connect the data output of the product(s) to the BIC-II communication port(s), starting with PORT1. Refer to
FIG. 3.
For multiple BIC-II applications, interconnect the BIC-II MillBus terminals. The MillBus cable shield should be
connected to a BIC-II at one end only, preferably the end where the MillBus Out connection is made. Refer to
FIG. 3G - MillBus / Multiple BIC-II Interconnection.
Connect the BIC-II to the host computer or PLC serial port. Refer to Appendices/PLC Requirements. Refer to
the computer or PLC manufacturers instruction manual to determine appropriate RS-232 or RS-422
connection requirements. Refer to FIG. 3H - Host Interconnection.
Connect the power wiring to the BIC-II(s). Refer to FIG. 3I - Power Wiring Detail.
All wiring must be done in conjunction with
approved conduit, boxes and fittings, and to procedures
in accordance with all governing regulations.
INTERNAL CHECKS
Ensure the BIC-II (board B) voltage selection jumper (J2) and MillBus common selection jumper (J1) are in
the appropriate position. Refer to FIG. 3I - Power Wiring Detail.
Ensure the 100 Ω termination resistor is installed on the proper unused BIC-II MillBus terminals. Refer to FIG.
3G - MillBus / Multiple BIC-II Interconnection.
Ensure jumper(s) are properly connected to the unused RS-232 or RS- 422 BIC-II terminals. Refer to FIG. 3H
BIC-II - Host Interconnection.
Do not operate the BIC-II with the ground (earth) wire disconnected.
■
PL-371 3 – 1