SELF TEST
This message instructs the BIC-II to perform an on-line self test and return the results.
Data ASCII Example
Field Character Description
SOM STX start of message
BIC-II I.D. 01 BIC-II #1
COMMAND B6 perform a self test
EOM CR end of message
The response to the host from the BIC-II will indicate the results of the self test and will consist of the following
data fields:
SOM BIC-II I.D. COMMAND STATUS EOM
If the self test was successful, the status field characters will be 000.
A Status field code of 128 indicates that the BIC-II power was interrupted during a Port Initialization. Perform
a Port Initialization Verification on each port of the BIC-II specified and transmit a new Port Initialization to the
PORT# in question.
Status field codes of any other value indicate BIC-II faults that cannot be corrected by the user.
ERROR RESET
This message resets the Status field error indication to 000. If an error reset is not performed the status field
error indication will remain unaltered, even if future self tests are successful.
Data ASCII Example
Field Character Description
SOM STX start of message
BIC-II I.D. 01 BIC-II #1
COMMAND B7 perform an error reset
MASK 000 reset all status field error bits
EOM CR end of message
ASCII characters 000 are used to reset the Status field indication to 000. Other values are reserved for
Milltronics service personnel.
PROGRAM
TIP
Upon a fault indication, performing a ‘Error Reset’
followed by another ‘Self Test’ is recommended to
ensure the fault indication was not due to
temporary transient noise. If the cause of any
persistent error cannot be determined contact
Milltronics or your distributor.
PL-371 8 – 9